Automatic, Discrete, Random Access, Bench-top
STAT sample priority
Throughput: Up to 20O tests/hour, up to 400 tests/hour with ISE (Optional)
Measuring principles: Absorbance photometry, Turbidimetry, lon Selective Electrode technology
Methodology: End-point, Fixed-time, Kinetic, optional ISE
Single/Dual/ reagent chemistries
monochromatic / bi-chromatic
Original system pack reagent ready to use
Close system and open system is optional
Reagent/Sample Handling
Reagent/Sample tray: 80 positions for reagents and 40 positions for samples in 24-hour refrigerated compartment (2~12 C )
Reagent volume: 10~250ul, step by 0.5uI
Sample volume: 2~45ul, step by 0.1ul
Reagent/Sample probe: Liquid level detection, vertical collision protection and inventory checking, reagent pre-warming
Probe cleaning: Automatic washing for interior and exterior
Carry over < 0.05%
Automatic sample dilution: Pre- dilution and post-dilution
Internal bar code reader (optional)
Capable to communicate with LIS in bi-directional mode
Reaction System:
Cuvette: 40 disposable cuvettesReaction volume: 100~360ul
Operating temperature: 37 C +0.1 C
ISE Module (optional)
Measuring K+, Na+, Cl-
Mixing Unit
Independent mixing bar
Optical System
Light Source: Halogen-tungsten lamp
Wavelength: I 8 wavelengths, 340nm, 405nm, 450nm, 510nm, 546nm, 578nm, 630nm,670nm
Absorption range: O~4.O Abs (10mm conversion), resolution 0.O001Abs
Stray Light: 5.6Abs
Control and Calibration
Calibration modes: Linear (one point, two points and multipoints), Logit-Log 4P, Logit-Log 5P, spline, exponential, polynomial, parabola
Control Rules: X-R, L-J, Westgard multi-rule, Cumulative sum check, twin plot
Operation Unit
Operation system: Windows 8
Interface: RS-232
Working Conditions
Power Supply: 200~240V, 50/60Hz, ≤1000VA or 100~130V, 60Hz, ≤10OOVA
Dimension: 690 mm (length) x58O mm (depth) x595 mm(height)
Weight: 47 kg
Water Consumption:< 2 L/H
Reviews