Product Specification
Class
Class II
Product Type
Type A2
Model
HR40
Operator
1
Interior Dimensions
1200*610 *680 mm
Overall dimensions
1360*780 *2200 mm
UV light
Greater Than 30w
Noise
Less than 62dB(A)
IIIumination
Greater Than 800Lx
Inflow Air Velocity
Greater Than 0.6m/s
Downflow Air Velocity
Greater Than 0.3+- 0.02m/s
Cleanliness
CLASS 100@ Greater Than 0.5um
Product
Other Details:
Product Feature: Fission structure : The front panels are detachable and easier to remove for filter and filter. Visible windows: with antiultraviolet toughened glass and the height of glass can be adjustable. If the height is higher than safe level, the audible and visual alarm will be opened.
The cabinet: all manufactured by 304??? stainless steel material and the left, right and back bins of cabinet are manufactured together with arc angle will make easier to clean the side of cabinet.
Cabinet feature: Contaminated Pressure Plenum: Negative Unitized drain pan beneath the work surface allows best arm position the user’s lap.
The airflow design creates a more impenetrable air barrier or momentum air curtain at the front of the cabinet, increasing the cabinet’s protective capabilities and avoiding cross infection.
This kind of design standard is comply with NSF4911 standard. With strong wind form the draught intake in side of cabinet, which will clear the round airflow up in lateral cabinet and prohibit the moving of contamination.
Ultra-high efficiency filter system: input and output of air will be filtrated by ULPA filter and its efficiency is more than 99.999%. the pressure sensor will watch the working condition of filter from time to time and alarm the operator if the operator if the pressure is over the setting value.
Ultraviolet light safety: the UV light cannot be opened when the fan, visible window and fluorescent light are running.
Interlock function for two fans ( blower and exhaust fan ):
When the exhaust fan stops, the blower will stop automatically and alarm device will begin to work. When the blower stops, the exhaust fan will continue to work and the alarm system will indicate the fault of blower.
Displaying data in the screen:Â Air-exhaust value ( m3/h ) the velocity of air laminar flow ( m/s ) counter for exhaust fan counter for blower.
Safety control system:Â Alarm for low air exhaust value
Alarm for low air-velocity of down laminar flow
Alarm for fault of exhaust fan
Alarm for closing condition of visible window
When the visible windows is opened, the UV light will be closed automatically.
When the fluorescent light is on working, the UV light will be closed automatically.
When the blower begin to work and then the UV light will be off after air laminar flow is created.
The blower will be off automatically if the exhaust fan happens some trouble.
The exhaust fan will increase his rotate speed automatically in order to keep the working space in negative pressure inside cabinet in case something wrong happens with the blower.
Microprocessor will counteract the gradually increased resistance in filter.
It will keep air inflow and downflow in normal condition under any conditions